Enable-IT 868 Instrukcja Użytkownika Strona 118

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 246
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 117
Using the Ultrasonic Thickness Gage 9-9
December 2003
Calibrating the Thickness Gage (cont.)
3. The TransPort displays one of the test block lengths. This is
the end you should use to calibrate first. Apply couplant to the
appropriate end of the test block and place the thickness
transducer on the block. The transducer face should be flat
against the surface of the test block.
4. The screen should display approximately the correct thickness
of the test block. Press the [F1] key to select “CAL.” The
TransPort locks in the calibration.
5. Repeat steps 3 and 4 for the other end of the test block.
6. Press [F2] (or [EXIT]) to exit.
The TransPort returns to the thickness gage display. You are now
ready to make pipe thickness measurements.
If you do not wish to make thickness measurements, press the
[EXIT] key until the TransPort returns to making flow
measurements.
Przeglądanie stron 117
1 2 ... 113 114 115 116 117 118 119 120 121 122 123 ... 245 246

Komentarze do niniejszej Instrukcji

Brak uwag